2-wire Digital Meter (Loop Powered) 6DV/6DV-B

4-digit, Input Loop Powered
- 70 mm dia., D 21 mm (6DV, 6DVI)
- W 110 x H 118 x D 92 mm (6DV-B, 6DVI-B)
- 4-digit red LED (-1999 to 9999)
- 4-20 mA input powered
- No external power source required
- Scaling & linearization selectable via the front control buttons
- Mountable on top of head-mount transmitter, installed together in an outdoor encloure (6DV, 6DVI)
- NEMA 4X, IP66 / IP67 (6DV-B, 6DVI-B)
Depending on the model.



-
Intrinsically safe approvals are available from ATEX Directive (Europe) and FM (US, Canada). Differences between the 6DV and the 6DVI exist only in safety parameters (See Table).
A safety barrier with matching safety parameters to the 6DV or 6DVI must be installed for complete intrinsic safety. The safety barrier should meet the Ex-data of this unit and must be approved for hazardous location.Table
Operating Temperature Temp Class Operating Temperature 6DV 6DVI FM/ATEX T4 Intrinsically safe T4 -40 to +80°C -40 to +85°C T5 -40 to +60°C -40 to +60°C Intrinsic Safety Ex Data 6DV 6DVI Ui (Vmax) 30V DC 28V DC Ii (Imax) 100mA DC 120mA DC*1 Pi (Pmax) 0.75W 0.84W Ci 1.0 nF 1.0 nF Li 0 mH 0 mH *1. Max. current allowed for the safety approval; Not to guarantee the product’s performance.
-
Related products
A3DYH 6BX-E 27HU Galvanic Isolators Outdoor Enclosure 2-wire HART
Temperature TransmitterMD6T MD6N MD6P Lightning Surge Protectors