Strain Gauge & Receiving Instrument
-
Lightning Surge Protector for Strain Gauge (ultra-slim)
-
- Ultra-thin 7-mm-wide module can be mounted in high density
- Excellent protection employing multi-stage SPD circuits
- DIN rail mounting and grounding
- Shield terminal provided
- Protects strain gauges and transmitters
- CE / ATEX for MD7LC-1 only
- Maximum continuous operating voltage (Uc):
±0.2 V (Out. side), 15 V (Exc. side)
- Voltage protection level (Up):
MD7LC-0: ±15 V @0.2 kV (Out. side), 30 V @0.2 kV (Exc. side)
MD7LC-1: ±40 V @6 kV (Out. side), 50 V @6 kV (Exc. side)
- Discharge current capacity:
MD7LC-0: 100 A (8/20 μs) (Out. side, Exc. side)
MD7LC-1: 20 kA (8/20 μs) (Out. side), 1.0 kA (10/350 μs) (Exc. side)
-
Lightning Surge Protector for Strain Gauge Use
-
- Designed to protect strain gauge transmitters from lightning surge damage that enters on the wiring between the strain gauge and the transmitter
- Absorbs surges only without affecting instrumentation signal
- No interruption of signal by unplugging the protector element
- Maximum continuous operating voltage (Uc):
Out. side: ±0.3 V
Exc. side: 15 V
- Voltage protection level (Up):
Out. side: ±15 V
Exc. side: 30 V
- Discharge current capacity:
Out. side: 100 A (8/20 μs), 25 A (10/1000 μs)
Exc. side: 50 A (8/20 μs), 25 A (10/1000 μs)
-
Lightning Surge Protector for Strain Gauge Use
-
- Designed to protect strain gauge transmitters from lightning surge damage that enters on the wiring between the strain gauge and the transmitter
- Absorbing surges only without affecting instrumentation signal
- Shallow depth
- DIN rail mounting
- Maximum continuous operating voltage (Uc):
Out. side: ±0.3 V
Exc. side: 15 V
- Voltage protection level (Up):
Out. side: ±15 V
Exc. side: 30 V
- Discharge current capacity:
Out. side: 100 A (8/20 μs)
Exc. side: 50 A (8/20 μs)
-
Lightning Surge Protector for Strain Gauge Use (6-wire remote sensing)
-
- Designed to protect strain gauge transmitters from lightning surge damage that enters on the wiring between the strain gauge and the transmitter
- Six-wire remote sensing
- Absorbing surges only without affecting instrumentation signal
- Maximum continuous operating voltage (Uc):
Signal side: ±0.3 V
Excitation side: 15 V
- Voltage protection level (Up):
Signal side: ±20 V
Excitation side: 140 V
- Discharge current capacity: 5000 A (8/20 μs)
- Response time: 4 nsec.
Self-Synch & Receiving Instrument
-
Lightning Surge Protector for Self-Synch (ultra-slim)
-
- Ultra-thin 7-mm-wide module can be mounted in high density
- Excellent protection employing multi-stage SPD circuits
- DIN rail mounting and grounding
- Shield terminal provided
- Protects self-synch transmitters and converters
- Maximum continuous operating voltage (Uc): ±130 V
- Voltage protection level (Up): ±450 V
- Discharge current capacity: 20 kA (8/20 μs), 1.0 kA (10/350 μs)
-
Lightning Surge Protector for Self-Synch Use
-
- Designed specifically for self-synchronizing motor circuit
- Protects self-synch transmitter and converter from damage caused by surges on the wiring
- Absorbs surges only without affecting instrumentation signal
- No interruption of signal by unplugging the protector element
- Maximum continuous operating voltage (Uc): A to B to C: ±170 V
- Voltage protection level (Up): 1 to 2 to 3: ±450 V
- Discharge current capacity: 5000 A (8/20 μs)